Figure 1. Morbidostat run procedure scheme.
- Calibration
- Morbidostat assembly
- Filling of the feed lines
- Program setup
- Innoculated tubes installation
- Start of the run
- Sampling
- Processing and storing samples
- Finish the experiment
- In-morbidostat washing after a run
- Disassembling and washing
The goal of calibration is to find slope (C) and intercept (V0) for equation OD600 = V0 + C*log(Voltage), where Voltage is output from the photosensors and OD600 is measured Optical Density (with 600nm wavelength) for a calibration standard.
We used BaCl2 suspension as calibration standards.
- Prepare:
- 150 ml 1% H2SO4
- 5 ml 1% BaCl2
- 6 culture tubes with screw caps identical with those used for reactors
- Prepare 25 ml of standards based on the table: McFarland
goal (McFarland) | 1% BaCl2 (uL) | 1% H2SO4 (ml) |
---|---|---|
0.25 | 63 | 24.93 |
0.5 | 125 | 24.88 |
1 | 250 | 24.75 |
2 | 500 | 24.50 |
3 | 750 | 24.25 |
4 | 1000 | 24.00 |
- Vortex standards vigorously.
- Take aliquots of standards to check McFarland and OD600 values on the McFarland densitometer and spectrophotometer.
- Mark true values on the upper part of the tubes.
- Set C=1 and V0=0.
- Measure values on warmed morbidostat and make linear regression TrueValue (Od or MF) ~ (Mesurement). The R^2 should be more then 0.97. You can find this linear regression with calibration.py [data.txt]
- Type values in the Calibration tab of Software.
- Check if all valves are open. If not – rotate step motors into the open position.
- Check if all wedges are open and allow free installation of the tubing into the valves. If not – use any thin instrument to move them (like a paper clip).
- Install pump 1 distribution line in the peristaltic pump. Don’t close the peristaltic pump.
- Install pump 1 distribution line in the valves.
- Do the same for the pump 2.
- Connect media connection lines. Pull peristaltic pump tubing to the position when stopper touch the pump case.
- Close peristaltic pumps.
- Install new air filters to the air entrance of the reactor cap tubing.
- Install reactors in the rack.
- Connect tubing, media distribution lines and air lines.
- Connect media bottles to the media connection lines.
- Add 20 ml of 5% sodium azide to the waste bottles.
- Connect waste lines to the reactors.
- Connect waste lines to the waste bottles.
- Turn on Morbidostat.
First fill line with a Pump 2 media (with a drug). Then Fill line with Pump 1 media and wash feed needle with no drug media.
- On the Connection Manager tab connect to the Morbidostat.
- On the Settings tab do check/un-check for all valves to be sure that valves are closed.
- On the same tab open valve 1 by checking corresponding On/Off check-box.
- On the same tab start pump 2 by checking Pump 2 On/Off check-box.
- Stop Pump 2 manually by un-checking Pump 2 On/Off check-box after media will fill the tubing.
- Close valve by un-checking corresponding On/Off check-box
- Repeat 3-6 for other reactors.
- In the Morbidostat Software on the Control tab check all tubes that are in experiment.
- On the Settings tab set for all reactors:
- Cycle time = 2 min
- Dilution time = 16 s
- Air pump = checked (On)
- Air Delay = -1 (No turning off of Air during pumping)
- For Cleaning = Pump1
- On the Control tab turn on OD Reading and Automatic Dilution.
- Wait for 2 washing cycles and click on Off button for Automatic Dilution and OD Reading at the end of the second cycle.
- On the Connection Manager tab connect to the Morbidostat and Agitation boards.
- On the Control tab:
- Check all tubes that are in the experiment.
- Set temperature (Standard = 37°C).
- While waiting for inoculated tubes turn on Temperature Control to warm optics.
- On the Settings tab:
- Set Lower Dilution thresholds (Standard = 0.15. Will be adjusted on the start).
- Set Upper Dilution thresholds (Standard = 0.3. Will be adjusted on the start).
- Cycle time (in minutes) (Standard = 15).
- Dilution time (in seconds) (Standard = 16 sec). 1sec = 0.3ml of added media.
- Laser strength should be the same as it was during calibration.
- Air Delay (Standard = 5 sec).
- Set For Cleaning = None.
- On the Message Logger tabs set log files for OD and Pump logs. Be sure that in the time settings menu “Include date” and “Include time” are checked.
- Close all lids.
- Put O-rings on the inoculated tubes.
- Open all morbidostat lids.
- Start from distant tubes. Rotate a reactor tube (NOT a cap) to unscrew it.
- Take off a cap. Be sure that needles do not touch anything. If necessary lift needles.
- Replace tube with inoculated one.
- Screw tube by rotating it (NOT a cap).
- Seal the O-ring on the rack and check that tube is touching the bottom of the rack.
- Repeat steps 4-7 for all reactors.
- Slightly burn all upper parts of the needles with a burner.
- Adjust needles:
- Set waste needle ~2 mm above the liquid surface.
- Set input needle tip under the surface ~15 mm lower than waste needle.
- Set sample needle above laser. Tip: Top of sample needle should be ~20mm above a top of the waste needle.
- Close all lids.
- On the Control tab turn on:
- OD Reading
- Temperature Control (if it was not already turned on)
- On the Settings tab turn on:
- Air.
- On the Agitation tab turn on:
- Agitation.
- Wait ~20 min to warm optics.
- Get current OD from the Optical Density plot and count it as zero OD.
- On the Settings tab adjust Lower Dilution and Upper Dilution thresholds by adding current OD.
- IMPORTANT: Turn on logging for OD and pumps on the Message Logger tabs.
- On the Control tab turn on:
- Automatic Dilutions
- Visually check system integrity (air bubbling, liquid levels, OD plots, temperature) after first dilution.
- Have ready sterile Pump 1 media for dilution at 37°C
- Prepare 6x15 mL falcons marked with sample names (containing reactor number and time point ID), 10 mL syringes
- Stop air, OD reading, dilution on software tab
- Take samples, one by one, from each reactor:
- Flame and lower the sampling needle; aspirate 10 mL in syringe, dispense to respective 15 mL tube;
- Take 10 mL Pump 1 media, inject into the same port
- Leave syringe in place, lift the needle, tape over the syringe (to avoid displacement)
- Repeat for all reactors that you want to sample
- Check all levels, turn on air, check bubbling
- Close box, turn on reading, check values;
- If OK, turn on dilution.
- Glycerol stocks:
- Prepare 6 sterile Eppendorf tubes, dispense 300 µL of 50% glycerol each
- To each – add 600 µL from each sample, mark on top 1B, 2B…. 6B or whatever letter corresponding to sampling cycle.
- Mix, spin down, freeze in liquid N2, put in -80 box
- Cell pellets (for genomic DNA):
- Spin remaining samples (~ 9.4 mL) in Beckman: RPM=4,000; Time = 7 min; Temp= 10°C
- Remove supernatant: pour off the majority and use 1 mL pipettor to aspirate remainder (alternatively use vacuum aspiration)
- Resuspend each pellet in 10 mL saline: Resuspend using 1 mL pipette to pull sample up and down until uniform and no chunks are present (vortex is much less effective)
- Spin again, same parameters
- Remove supernatant, resuspend in 5 mL each (washing step, optional)
- Variation – make one or two duplicate aliquots.
- Spin, resuspend by pipette in 1.5 mL (for one aliquot); [option - in 3 mL (to split between two tubes, depends on the number of cells and importance)]
- Transfer to color-coded (by cycle) 2 mL Eppendorf, label on top
- Spin in microcentrifuge (5 min, 14,000)
- Remove super, freeze pellets in in liquid N2, put in -80 box
- On the Settings tab turn off:
- Air
- On the Control tab turn off:
- OD Reading
- Automatic Dilutions
- Temperature Control
- Take last samples (Look in the sampling protocol).
- On the Agitation turn off:
- Agitation
- Prepare
- ~300 ml of 10% bleach in a bottle.
- ~300 ml of 20% ethanol in a bottle.
- Two tubing ~30 cm with female luer connectors
- Disconnect media bottles, save residual media if needed and bring it for washing.
- Disconnect media connection tubing part before both peristaltic pumps.
- Connect the bottle with the fresh bleach to both peristaltic pumps with any tubing. Close all morbidostat lids to decrease bleach odor.
- In the Morbidostat Software on the Settings tab set for all reactors:
- Cycle time = 2 min
- Dilution time = 16 s
- Air pump = checked (On)
- Air Delay = -1 (No turning off of Air during pumping)
- For Cleaning = Pump1
- In the Morbidostat Software on the Control tab check all the tubes and set:
- OD Reading = ON
- Automatic Dilutions = ON
- Wait 5 minutes.
- Set Automatic Dilutions = OFF. Wait until program is finished (The green indicator near Automatic Dilution become blank).
- In the Morbidostat Software on the Settings tab set For Cleaning = Pump2
- Redo steps 6-8.
- Wait 10 minutes.
- Switch the bleach bottle with the ethanol bottle.
- Redo steps 5-11.
- Disconnect ethanol bottle and leave pumps tubing disconnected.
- Redo steps 5-11 to empty feed lines.
- Turn off OD reading and Air in the Software and then turn off Morbidostat power switch.
- Prepare:
- The plastic tray.
- 50ml falcon with 10% bleach (Use the leftovers from the bleach bottle).
- Magnetic stick retriever.
- Culture tube holder.
- Disassemble waste lines from reactors. Rinse waste lines needles in the bleach. Put them in the tray.
- Disconnect all feed distribution lines and air line. Discard air filters.
- One by one for each reactor:
- Unscrew reactor caps and put it in the tray.
- Pour out all liquid from the tube. With the help of the magnetic stick retrieve - keep stirring bar inside the tube while pouring.
- Put the tube on the holder.
- Put the tray under feed lines ends.
- Open all valves.
- Pull out all tubing out of valve holders and put them in the tray.
- Open peristaltic pumps.
- Pull out the tubing out of the peristaltic pump.
- Bring tubing and tubes for washing.
- Bring waste bottles to the hood.
- Pour bleach to waste bottles in a volume about 10% of waste.
- Close Waste bottles to minimize an odor.
- Let waste stay for ~30 min.
- Pour out waste.
- Bring waste bottles to the washing.
- Wash everything with soap water and laboratory brush. Then rinse thoroughly with DI-water.
- Let everything dry. Water from tubing could be sucked with vacuum.
- Assemble reactors and close sample needle with a stopper.
- Foil all open connectors and pack everything for autoclaving. NOTE: Don’t forget to put stirring bars to reactors and media bottles prior to autoclaving.
- Autoclave 30 min on the gravity cycle.