Issue:
When suspending a Medtronic pump FreeAPSX recognises and reports a red loop.
When un-suspending using the pump the pump will resume on either its temp or normal basal rate but FreeAPSX will not move from a red loop state showing message: "Error: Invalid Pump State: Pump Suspended:
Additional diagnostic:
RL connected over BT at -53dB
Data from un-suspend passes to FreeAPSX and displayed in basal graph correctly.
App forced restart has no effect
iPhone restart no effect
RL restart no effect
Treatment data from manual bolus passed to treatment log and displayed on graph successfully.
Workaround:
Re-suspend pump, clear and 0 basal and enable "Unsuspend if No Temp"
Pump and FreeAPSX will then unsuspend successfully.
The text was updated successfully, but these errors were encountered:
Issue:
When suspending a Medtronic pump FreeAPSX recognises and reports a red loop.
When un-suspending using the pump the pump will resume on either its temp or normal basal rate but FreeAPSX will not move from a red loop state showing message: "Error: Invalid Pump State: Pump Suspended:
Device:
FreeAPSX 0.2.5
Medtronic 772 WW
Rileylink: Firmware Emalink for RL CC1110 V2.219 / BLE113 V3.11
iPhone 13 Pro: iOS 15.1
Additional diagnostic:
RL connected over BT at -53dB
Data from un-suspend passes to FreeAPSX and displayed in basal graph correctly.
App forced restart has no effect
iPhone restart no effect
RL restart no effect
Treatment data from manual bolus passed to treatment log and displayed on graph successfully.
Workaround:
Re-suspend pump, clear and 0 basal and enable "Unsuspend if No Temp"
Pump and FreeAPSX will then unsuspend successfully.