Issue:
When suspending a Medtronic pump FreeAPSX recognises and reports a red loop.
When un-suspending using the pump the pump will resume on either its temp or normal basal rate but FreeAPSX will not move from a red loop state showing message: "Error: Invalid Pump State: Pump Suspended:

Device:
FreeAPSX 0.2.5
Medtronic 772 WW
Rileylink: Firmware Emalink for RL CC1110 V2.219 / BLE113 V3.11
iPhone 13 Pro: iOS 15.1

Additional diagnostic:
RL connected over BT at -53dB
Data from un-suspend passes to FreeAPSX and displayed in basal graph correctly.
App forced restart has no effect
iPhone restart no effect
RL restart no effect
Treatment data from manual bolus passed to treatment log and displayed on graph successfully.

Workaround:
Re-suspend pump, clear and 0 basal and enable "Unsuspend if No Temp"
Pump and FreeAPSX will then unsuspend successfully.

@jonny772
Copy link
Author

jonny772 commented Jan 2, 2022

3CE8E298-77C4-43E6-B8B8-BEBF1DA7FB81
8BECC032-160D-45F7-A975-7D6FBA61F8D6
385DBDA5-A1FE-4B66-B150-FE0FD35F787B
8BCDEE21-DD53-4D99-B122-A499FF0EC6BE

@Typ1er
Copy link

Typ1er commented Jan 6, 2022

I call up the RileyLink menu and continue the delivery.
IMG_95074A7A64EE-1