Notifications
You must be signed in to change notification settings - Fork 5
Room
These are similar to the mesoSPIM-V5, except the optical table requirement.
-
The floor should be solid concrete to avoid vibrations from walking.
-
Windows should be covered with black curtains or absent.
-
Enough space for a desk with a large (40inch) screen for the imaging computer and 2-3 chairs.
-
Power outlets: Foresee enough power outlets for the instrument. Total power uptake is a few kW at maximum
-
Optical table is not required. The Benchtop mesoSPIM, as its name suggests, can be placed on a sturdy bench, desk, or even a cart.
-
Active ventilation is required when working with BABB, DBE, or any other toxic chemicals.
-
Air conditioning, especially for long-running acquisitions.
-
For sample preparation, a wet lab bench, a hood and a sink in the vicinity of the microscope.
-
Chemicals must be handled in a fume hood, and stored in a safe cabinet (many of them are toxic and/or highly flammable).
-
Observe all relevant chemical, laser, trash disposal, and fire hazard safety rules of your institute. Take the safety rules seriously, they are written in blood.
-
Do not dispose chemicals into the water stream.
- Overview
- Room requirements
- Safety
- Parts list
-
Assembly
- Overview
- Detection arm
- Excitation arms
- Sample stages
- Electronics
- Lasers
- Immersion chambers (TODO)
- Front cover (with webcam)
- Enclosure (TODO)
- Usage
- Troubleshooting and typical errors (TODO)
- Maintenance(TODO)
- Technical notes